Simco Fmx002 Manual Full [upd] Jun 2026
Features a dedicated data port to transmit measurements to chart recorders or data loggers.
Pull the meter away immediately to protect the sensor plate circuitry. Fluctuating or Unstable Readings Cause: Missing ground connection or high ambient humidity.
Only qualified personnel should operate or service the equipment to avoid injury.
Typically features two ranges: 0 to ±2 kV and 0 to ±20 kV, allowing for precise readings of both low and high-level static.
Solution: Securely connect the grounding wire to earth ground. Replace the battery if it is low. Cause: Dead battery or internal hardware failure. Solution: Change the 9V battery immediately. simco fmx002 manual full
Are you trying to solve a right now?
Comprehensive Guide to the Simco FMX-002 Electrostatic Fieldmeter: Full Manual and Operational Instructions The Simco FMX-002 Go to product viewer dialog for this item.
Understanding the technical boundaries of the FMX-002 ensures accurate data collection and prevents internal circuit damage.
Built-in converging lights ensure the meter is held at the exact 1-inch (25mm) measuring distance. Features a dedicated data port to transmit measurements
Precise execution of the following steps is critical to prevent inaccurate data collection caused by stray ambient charges. Step 1: Grounding the Device and Operator
Displays positive (red) or negative (blue) charge polarities.
Point the sensor plate away from any charged bodies or static sources. Press the gray button. Verify that the digital readout displays exactly 0.00 . Step 4: Positioning and Measurement Aim the sensor at the target object surface.
0 to ±1.49 kV (Low range); ±1.0 kV to ±22.0 kV (High range). Accuracy: ±10% of reading. Measuring Distance: 25 mm ±0.5 mm (1 inch). Response Time: Less than 1 second. Only qualified personnel should operate or service the
A mechanical rig that locks the FMX-002 precisely 1.00 inch away from the calibration plate. Calibration Process
If it does not read zero, press the button. The meter will beep and reset its baseline. Step 3: Taking a Measurement Point the sensor toward the target object.
Here's a step-by-step guide to help you get started with the Simco FMX002:
Turn on the device by pressing the button once. The unit will beep.