The manual includes SQL query templates for extracting raw data directly into statistical software like Minitab or JMP for process control charting.
What sets Syngistix apart is its cross-platform design. The software offers a harmonized user experience across PerkinElmer's AA, ICP, and ICP-MS platforms, featuring a unique icon-based design that simplifies navigation and walks users through every analysis—from setup to data acquisition to reporting results. The interface mirrors your analytical workflow with a logical left-to-right arrangement of steps, making it intuitive even for novice users.
[Select Element] │ ▼ [Recommended Specs] ──(Optional Modification)──► [Set Wavelength & Slit] │ ▼ [Integration/Delay] │ ▼ [Calibration Profile] Step-by-Step Method Creation Product: Syngistix for AA Software - PerkinElmer
Not all peaks are perfect. The manual teaches you how to manually integrate (re-measure peak area or height) asymmetric furnace peaks caused by interfering matrices. You can adjust start and end points, and the software automatically recalculates the concentration. syngistix for aa software manual
Flow Injection Atomic Spectroscopy (FIAS), Mercury Hydride Systems (MHS), and the Flow Injection Mercury System (FIMS).
: Allows selection of the running technique, such as Flame, Graphite Furnace, or specialized methods like FIMS (Flow Injection Mercury System). Core Procedures 1. Method Setup
| Symptom | Likely cause | Action | |---------|--------------|--------| | No signal | Lamp not lit or misaligned | Check lamp current, Energy > Optimize | | High noise | Dirty burner or flame instability | Clean burner, check gas flows, nebulizer | | Poor R² on curve | Wrong range or contamination | Remake standards, check blank, use quadratic fit | | Furnace peak tailing | Incorrect atomization temp | Increase atomization step by 100–200°C | | Communication error | USB cable or driver | Reconnect, restart Syngistix, reboot PC | | QC fails repeatedly | Standard degradation or matrix effect | Fresh stds, use matrix matching, check autosampler probe | The manual includes SQL query templates for extracting
, it replaced the older WinLab32 software to provide a more intuitive, icon-based experience for elemental analysis. Core Software Versions Syngistix for AA
For power users, Syngistix supports numerous keyboard shortcuts to speed common operations. Navigation keys allow movement through sample lists, and specialized shortcuts are available in Analytic mode for functions like adding read delay points.
500 GB HDD/SSD with at least 20 GB of free space for database storage. The interface mirrors your analytical workflow with a
The Automated Analysis Control window displays the sequence of methods and samples to be analyzed. You can:
Syngistix offers features that go beyond simple data acquisition, allowing for improved laboratory efficiency.